User Contributed MET/CAL PROCEDURE ============================================================================== INSTRUMENT: FLUKE 8300A DATE: 22-Apr-87 AUTHOR: User Contributed REVISION: V 4.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 34 NUMBER OF LINES: 78 CONFIGURATION: Fluke 5100B Fluke 8506A CONFIGURATION: ============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP Check for any obvious physical damage. 1.001 DISP Check that all moving parts operate firmly and smoothly 1.001 DISP Check for proper fuse size. 1.001 DISP Energize the T.I. and allow 30 min. warm-up. 1.002 HEAD DC VOLTAGE TESTS 1.003 5100 10 1V 0.04% 2W 2.001 5100 10 10V 0.013% E 2W 3.001 5100 100 10V 0.04% 2W 4.001 5100 100 100V 0.013% E 2W 5.001 5100 1000 100V 0.04% 2W 6.001 5205 1000 1000V .013% E 2W 7.001 5100 10 -1V 0.04% 2W 8.001 5100 10 -10V 0.013% E 2W 9.001 5100 100 -10V 0.04% 2W 10.001 5100 A 0V 1U 2W 11.001 5100 A 9V 0.04% 2W 12.001 EVAL Does the T.I. up-range properly? 13.001 OPBR Does the T.I. have the AC option? 13.002 JMPF 21.001 13.003 HEAD AC VOLTAGE TESTS 13.004 DISP Set T.I. to AC on the 1000 range and short input 13.004 DISP terminals. 13.004 DISP Adjust the AC ZERO control on the front panel for 13.004 DISP a display of 000.00 +/- .30 13.005 DISP Remove short. 13.006 5100 1 1V 0.13% 400H E 2W 14.001 5100 10 10V 0.13% 400H E 2W 15.001 5100 100 100V 0.13% 400H E 2W 16.001 5100 1000 100V 0.4% 400H 2W 17.001 5100 1 1V 0.13% 10KH E 2W 18.001 5100 10 10V 0.13% 10KH E 2W 19.001 5100 100 100V 0.13% 10KH E 2W 20.001 5100 1000 100V 0.4% 10KH 2W 21.001 OPBR Does T.I. have millivolts option? 21.002 JMPF 21.003 HEAD MILLIVOLTS DC TESTS 21.004 DISP Set T.I. to MVDC and 100 range. 21.004 DISP Short input terminals and adjust MV ZERO for a 21.004 DISP display of 00.000 +/-.003. 21.005 5100 100 10mV 0.07% 2W 22.001 5100 100 100mV 0.016% E 2W 23.001 5100 1000 100mV 0.04% 2W 24.001 5100 1000 1000mV 0.13% 2W 25.001 5100 100 -10mV 0.07% 2W 26.001 5100 100 -100mV 0.016% E 2W 27.001 5100 1000 -100mV 0.04% 2W 28.001 5100 1000 -1000mV 0.013% E 2W 29.001 HEAD RESISTANCE TESTS 29.002 RESF 1 1KZ 0.013% 2W 30.001 RESF 10 10KZ 0.013% 2W 31.001 RESF 100 100KZ 0.013% 2W 32.001 RESF 1000 1MZ 0.013% 2W 33.001 RESF 10 10MZ 0.08% 2W 34.001 DISP Verification procedure complete. 34.001 DISP Perform any and all Preventive Maintenance and check 34.001 DISP off appropriate boxes on the TECCAM card. END